Caprac-t Wipe Test/Well Counter
No need to worry about meeting the regulations in 10 CFR Part 35.315 (200 dpm requirements for unrestricted areas and iodine contamination). The Capintec™ CAPRAC-t Wipe-Test Counter handles all of them.
Brand | |
---|---|
Model | CAPRAC -t Wipe Test / Well Counter |
No need to worry about meeting the regulations in 10 CFR Part 35.315 (200 dpm requirements for unrestricted areas and iodine contamination). The Capintec™ CAPRAC-t Wipe-Test Counter handles all of them.
Measure for measure, no other well counter offers the speed, accuracy and complete range of built-in features provided by the compact CAPRAC-t counter. It performs a wipe test in just six seconds (for 1 nCi) and detects extremely low levels of activity with the accuracy only a NaI drilled-well detector can provide.
The CAPRAC-t system can also serve as a single-well gamma counter in departments that do not need multi-sample changers. User-defined protocols, trigger levels, and counting times are a “snap”. A 256-channel pulse-height analyzer permits built-in gamma spectroscopy. Definable conversion factors for specific radionuclides allow the CAPRAC-t counter to calculate results in cpm, dpm, nCi, cps, dps, or kBq. The CAPRAC-t unit also displays photon-energy histograms.
The CAPRAC-t counter is engineered and built for years of consistent, reliable performance and is curie or becquerel selectable. The unit has automatic background subtraction and self-diagnosis programs for systems testing with optimized signal-to-noise ratio. It includes a lead outer shield (1.3 cm thick) with optional auxiliary shield available.
Features
- 8 in SVGA touch screen colour display
- 256 channel MCA with detailed spectrum for identification analysis
- NaI drilled-well crystal detector
- Automatic energy calibration, constancy check, and background subtraction
- Automated well QC including chi-square and MDA
- Manual and Automatic ROI
- Sets user-definable protocols for wipes
- Printouts of all data for permanent records with the optional printer
- Direct readouts with spectrum display are in cpm, dpm, curies and bequerels
- Meets all state and NRC wipe test requirements
- USB and RS-232 communication ports for PC and printer
- Compatible with nuclear medicine information management systems via USB
- Lab tests include Schilling, plasma and red blood cell (RBC) volume
- Built-in database for test and wipe results as well as QC
Specifications
Performance:
- Type: Drilled-well crystal, NaI(Tl) scintillator
- Crystal Dimensions: 3.8 x 4.4 cm (1.5 x 1.75 in.)
- Shielding: 1.3 cm (0.5 in.) lead
- Shielding: 2.75 m (9 ft) interconnecting cable
- Channels: 256
- Counting Rate: 100,000 cps, max
Well Detector:
- Type: 1 1/2 in Sodium Iodide (NaI) drilled-well crystal detector
- 256 Channel MCA, manual and automatic ROI
- Trigger Levels: User-definable
- Automated Calibration and Background Subtract
- QC Tests: Reproducibility, Chi-square, MDA
Display Screen:
- Type: 8 in. VGA LCD color touch screen display
- Bq/Ci Reading: User selectable or fixed
- Activity Display: Selected radionuclide, efficiency, measured activity and display units (Bq/Ci)
- Count Rate Values: Wipe and lab test results
Standard Source Data:
- System Memory: Cs-137, Eu-152 standard sources
- Efficiency for commonly used isotopes
PC Port:
- Interface: RS-232 and USB
- Compatibility: Standard Nuclear Medicine Management Systems
Printer:
- Interface: RS-232 and USB
- Power Requirements:
- 100-240 V ac (50/60 Hz) 90 MA
Tests:
- Diagnostics: Full test of program, system memories
Measurements
Console Dimensions:
- Width: 23 cm (9.0 in.)
- Depth: 27 cm (10.5 in.)
- Weight: 3.4 kg (7.5 lb)
Chamber Dimensions:
- Height: 23.8 cm (9.38 in)
- Diameter: 15.2 cm (6 in)
- Weight: 6.9 kg (15.2 lb)
- Well Diameter: 1.7 cm (.67 in.)
- Well Depth: 3.8 cm (1.5 in.)
- Cable Length: 2.7 m (9 ft)
Cables:
- Power: 1.8 m (6 ft)
- Printer1: 1.8 m (6 ft)