Caprac-t Wipe Test/Well Counter

No need to worry about meeting the regulations in 10 CFR Part 35.315 (200 dpm requirements for unrestricted areas and iodine contamination). The Capintec™ CAPRAC-t Wipe-Test Counter handles all of them.

Brand

Model

CAPRAC -t Wipe Test / Well Counter

Add to Quote

No need to worry about meeting the regulations in 10 CFR Part 35.315 (200 dpm requirements for unrestricted areas and iodine contamination). The Capintec™ CAPRAC-t Wipe-Test Counter handles all of them.

Measure for measure, no other well counter offers the speed, accuracy and complete range of built-in features provided by the compact CAPRAC-t counter. It performs a wipe test in just six seconds (for 1 nCi) and detects extremely low levels of activity with the accuracy only a NaI drilled-well detector can provide.

The CAPRAC-t system can also serve as a single-well gamma counter in departments that do not need multi-sample changers. User-defined protocols, trigger levels, and counting times are a “snap”. A 256-channel pulse-height analyzer permits built-in gamma spectroscopy. Definable conversion factors for specific radionuclides allow the CAPRAC-t counter to calculate results in cpm, dpm, nCi, cps, dps, or kBq. The CAPRAC-t unit also displays photon-energy histograms.

The CAPRAC-t counter is engineered and built for years of consistent, reliable performance and is curie or becquerel selectable. The unit has automatic background subtraction and self-diagnosis programs for systems testing with optimized signal-to-noise ratio. It includes a lead outer shield (1.3 cm thick) with optional auxiliary shield available.

Features

  • 8 in SVGA touch screen colour display
  • 256 channel MCA with detailed spectrum for identification analysis
  • NaI drilled-well crystal detector
  • Automatic energy calibration, constancy check, and background subtraction
  • Automated well QC including chi-square and MDA
  • Manual and Automatic ROI
  • Sets user-definable protocols for wipes
  • Printouts of all data for permanent records with the optional printer
  • Direct readouts with spectrum display are in cpm, dpm, curies and bequerels
  • Meets all state and NRC wipe test requirements
  • USB and RS-232 communication ports for PC and printer
  • Compatible with nuclear medicine information management systems via USB
  • Lab tests include Schilling, plasma and red blood cell (RBC) volume
  • Built-in database for test and wipe results as well as QC

Specifications

Performance:

  • Type: Drilled-well crystal, NaI(Tl) scintillator
  • Crystal Dimensions: 3.8 x 4.4 cm (1.5 x 1.75 in.)
  • Shielding: 1.3 cm (0.5 in.) lead
  • Shielding: 2.75 m (9 ft) interconnecting cable
  • Channels: 256
  • Counting Rate: 100,000 cps, max

Well Detector:

  • Type: 1 1/2 in Sodium Iodide (NaI) drilled-well crystal detector
  • 256 Channel MCA, manual and automatic ROI
  • Trigger Levels: User-definable
  • Automated Calibration and Background Subtract
  • QC Tests: Reproducibility, Chi-square, MDA

Display Screen:

  • Type: 8 in. VGA LCD color touch screen display
  • Bq/Ci Reading: User selectable or fixed
  • Activity Display: Selected radionuclide, efficiency, measured activity and display units (Bq/Ci)
  • Count Rate Values: Wipe and lab test results

Standard Source Data:

  • System Memory: Cs-137, Eu-152 standard sources
  • Efficiency for commonly used isotopes

PC Port:

  • Interface: RS-232 and USB
  • Compatibility: Standard Nuclear Medicine Management Systems

Printer:

  • Interface: RS-232 and USB
  • Power Requirements:
  • 100-240 V ac (50/60 Hz) 90 MA

Tests:

  • Diagnostics: Full test of program, system memories

Measurements

Console Dimensions:

  • Width: 23 cm (9.0 in.)
  • Depth: 27 cm (10.5 in.)
  • Weight: 3.4 kg (7.5 lb)

Chamber Dimensions:

  • Height: 23.8 cm (9.38 in)
  • Diameter: 15.2 cm (6 in)
  • Weight: 6.9 kg (15.2 lb)
  • Well Diameter: 1.7 cm (.67 in.)
  • Well Depth: 3.8 cm (1.5 in.)
  • Cable Length: 2.7 m (9 ft)

Cables:

  • Power: 1.8 m (6 ft)
  • Printer1: 1.8 m (6 ft)