Radiation Alert Monitor 4
The Monitor 4 is compact, analog radiation detector. A general purpose survey meter capable of detecting alpha, beta, gamma, and x-rays over 3 selectable ranges. Proven and reliable over decades of use in the industry.
Brand | |
---|---|
Model | Monitor 4 |
The Monitor 4 is compact, analog radiation detector. A general purpose survey meter capable of detecting alpha, beta, gamma, and x-rays over 3 selectable ranges.
Proven and reliable over decades of use in the industry, the Monitor 4 has evolved into one of the best analog radiation detectors of today. Simple and ergonomic , a red count light flashes and a beep sounds with each count detected. Battery check and silent operation with the flick of your thumb.
Detector
Halogen-quenched uncompensated GM tube with thin mica window 1.5-2.0 mg/cm2 thick.
Effective diameter of window is 9.1 mm (.36 in.).
Operating Range
0-.5, 0-5, 0-50 mR/hr
0-500, 0-5,000, 0-50,000 CPM
or 0-5, 0-50, 0-500 μSv/hr (SI Scale Meter Option)
Range Switch
X1, X10, X100, Battery Check.
Energy Sensitivity
1000 CPM/mR/hr (Cs137).
Detects alpha down to 2.5 MeV; typical detection efficiency at 3.6 MeV is greater than 80%. Detects beta at 50 keV with typical 35% detection efficiency and beta at 150 keV with typical 75% detection efficiency. Detects gamma and x-rays down to 10 keV typical through the window, 40 keV minimum through the case.Monitor 4 Energy Response
Accuracy
Reference to Cs137
Typically ±15% from factory, ±10% with NIST Source Calibration NIST Calibration Additional
Display
Analog Meter holds full scale in fields as high as 100X maximum reading. CPM & mR/hr scale. Optional SI Scale Meter Available
Weight M4: 200 g (7.1 oz) M4EC: 224 g (7.9 oz) Size 209 x 71 x 50 mm (8.2 x 2.8 x 1.9 in.)
Features
Specifications
Detector
Halogen-quenched uncompensated GM tube with thin mica window 1.5-2.0 mg/cm2 thick.
Effective diameter of window is 9.1 mm (.36 in.).
Operating Range
0-.5, 0-5, 0-50 mR/hr
0-500, 0-5,000, 0-50,000 CPM
or 0-5, 0-50, 0-500 μSv/hr (SI Scale Meter Option)
Range Switch
X1, X10, X100, Battery Check.
Energy Sensitivity
1000 CPM/mR/hr (Cs137).
Detects alpha down to 2.5 MeV; typical detection efficiency at 3.6 MeV is greater than 80%. Detects beta at 50 keV with typical 35% detection efficiency and beta at 150 keV with typical 75% detection efficiency. Detects gamma and x-rays down to 10 keV typical through the window, 40 keV minimum through the case.Monitor 4 Energy Response
Accuracy
Reference to Cs137
Typically ±15% from factory, ±10% with NIST Source Calibration NIST Calibration Additional
Display
Analog Meter holds full scale in fields as high as 100X maximum reading. CPM & mR/hr scale. Optional SI Scale Meter Available
Measurements
Weight M4: 200 g (7.1 oz) M4EC: 224 g (7.9 oz) Size 209 x 71 x 50 mm (8.2 x 2.8 x 1.9 in.)