Pro-RF Res 150 (CBCT/DVT)

The phantom for acceptance and constancy testing of Cone-Beam CT / Volume Tomography (DVT) in fluoroscopy according to DIN 6868-150.

Brand

Model

Pro-RF Res 150

Code

PRO02-120

Add to Quote

The phantom for acceptance and constancy testing of Cone-Beam CT / Volume Tomography (DVT) in fluoroscopy according to DIN 6868-150.

 

Features

Complies with:

  • DIN 6868 – 150
  • CE certified
  • the manual provides detailed guidelines for carrying out each test, results assessment and registration

Technical Data

Technical Data (can be modified to customer specifications):

  • Dimensions: 120 x 120 mm 10 mm thick test module containing sets of cylindrical holes
  • Parallel to Z-Axis, of 0.5, 0.6, 0.7, 0.8, 0.9, 1.0 and 1.3 mm diameters
  • Two additional 25 mm PMMA plates

Measurements

Dimensions: 120 x 120 mm 10 mm