Pro-RF CDR (Radiographic)

The phantom for evaluation of imaging performance of radiographic systems. It allows wide-range low-contrast and high-contrast resolution evaluation and sensitometric measurements.

Brand

Model

Pro-RF CDR

Code

PRO02-402

Add to Quote

The phantom for evaluation of imaging performance of radiographic systems. It allows wide-range low-contrast and high-contrast resolution evaluation and sensitometric measurements.

Available in standard layout plus any custom configuration.

 

Features

Complies with:

  • IEC 61223-3-1
  • IPEM Report 91 (2005): Recommended Standards for the Routine Performance Testing of Diagnostic X-Ray Imaging Systems published by the Institute of Physics and Engineering in Medicine
  • CE certified
  • the manual provides detailed guidelines for carrying out each test, results assessment and registration

Technical Data

Technical Data (can be modified to customer specifications):

  • Diameter: 180 mm
  • Thickness: 10 mm
  • 3 circular rims on the outside of the phantom (Ø: 150, 160 and 170 mm) for geometry and collimation check
  • 17 low-contrast objects, 11 mm in diameter, producing contrast values from 0.2 to 7.5%
  • 17 small high-contrast objects, 0.5 mm in diameter, producing contrast values from 3.9 to 95.4%
  • 10 sensitometric measurement point details, 5.6 in diameter, producing contrast values from 0.11 to 1.0
  • Pattern for line pair resolution evaluation (from 0.5 to 14.3 LP/mm)
  • 1.0 and 0.5 mm high purity Cu attenuation filters (fits to the diaphragm housing)
  • Compact carrying case

Measurements

Diameter: 180 mm

Thickness: 10 mm