Pro-NM Performance

The phantom for performance evaluation of small animal systems or ultra-high resolution ECT systems.

Brand

Model

Pro-NM Performance

Code

PRO08-101 – standard version, PRO08-103 – version with the PET Lid

Add to Quote

The phantom for performance evaluation of small animal systems or ultra-high resolution ECT systems. It can be used to evaluate: center-of-rotation error, non-uniformity artifacts and spatial resolution. Available in two sizes for FOV at least 45 mm or 75 mm.

 

Features

Complies with:

– NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
– AAPM Report No. 9 – Computer Aided Scintillation Camera Acceptance Testing
– AAPM Report No. 22 – Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
– ACR–SNM (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticals
– CE certified
– the manual provides detailed guidelines for carrying out each test, results assessment and registration

 

Specifications

Technical data (can be modified to customer specifications):

at least 75 mm FOV version:

main cylinder:

– outside cylinder diameter: 80 mm
– inside cylinder diameter: 74 mm
– inside cylinder height: 65 mm

cold rods insert:
– rod diameters: 1.2, 1.6, 2.4, 3.2, 4.0 and 4.8 mm
– height of rods: 34 mm

At least 45 mm FOV version:

main cylinder:

– outside cylinder diameter: 50 mm
– inside cylinder diameter: 45 mm
– inside cylinder height: 65 mm

cold rods insert:
– rod diameters: 1.2, 1.6, 2.4, 3.2, 4.0 and 4.8 mm
– height of rods: 34 mm
optional carrying case