This stability (agar) phantom consists of a cylindrical phantom and agar material inside.
Using this phantom a Signal to Noise Ratio, Signal Fluctuation to Noise Ratio, drift, and other imaging measures over a 100-volume or 200-volume fMRI scan can be performed.
The agar phantom has characteristics similar to the T2 measures of a human head, but provides no change in signal. The T1 and T2 characteristics of the agar phantom at 3T are ~900 ms T1 and 30 ms T2.
complies with:
– ACR MRI Accreditation Program
– IPEM Report 80 “Quality Control in MRI”, 1998
– AAPM Rep. 28 “Quality Assurance methods & phantoms for MRI”, 1990
– AAPM Rep. 34 “Acceptance testing of MRI systems”, 1992
– AAPM Rep. 100 “Acceptance & Quality Assurance procedures for MRI facilities”, 2010
this is not an ACR approved phantom for accreditation
CE certified
the manual provides detailed guidelines for carrying out each test, results assessment and registration
Technical data (can be modified to customer specifications):
overall cylinder dimensions: 140 mm in diameter, 150 mm in height
cylinder made of PMMA
filled with agar gel with T1 and T2 characteristic at 3T of ~900 ms T1 and 30 ms T2
optional hard-sided carrying case (MR unsafe) (09-102)
Features
complies with:
– ACR MRI Accreditation Program
– IPEM Report 80 “Quality Control in MRI”, 1998
– AAPM Rep. 28 “Quality Assurance methods & phantoms for MRI”, 1990
– AAPM Rep. 34 “Acceptance testing of MRI systems”, 1992
– AAPM Rep. 100 “Acceptance & Quality Assurance procedures for MRI facilities”, 2010
this is not an ACR approved phantom for accreditation
CE certified
the manual provides detailed guidelines for carrying out each test, results assessment and registration
Specifications
Technical data (can be modified to customer specifications):
overall cylinder dimensions: 140 mm in diameter, 150 mm in height
cylinder made of PMMA
filled with agar gel with T1 and T2 characteristic at 3T of ~900 ms T1 and 30 ms T2
optional hard-sided carrying case (MR unsafe) (09-102)